RAGL Rotameter Flow Meter
Brand: YokogawaThe RAGL Laboratory Rotameter is ideal for measuring low flows of liquids and gases, designed with a conical glass metering tube that has a free rotating float, which is mounted in a vertical pipeline with flow direction upwards.
In addition, its anti-static metering tubes allow for measurement of small quantities of gas, delivering high accuracy at low flow rates. The flow is indicated by the top of the float and can be read from the standard scale on the metering tube or from a connected scale.
Meets High Safety Standards
The RAGL’s tube is sealed with o-rings and additional tube gaskets, which are fixed with pressure flanges for high tightness, providing a higher level of safety.
Features- Needle valves direct assembled in the head of the inlet or the outlet
- Inductive ring sensors as limit switches available
- Inlet and outlet flow controller available
- Frame for panel mounting available
- Foot plate for table installation available
Specifications
Brand
Accuracy
- ± 1.6% of Full Scale
- ± 2.5% of Full Scale
- ± 4% of Full Scale
- ± 6% of Full Scale
Approvals & Certifications
- ATEX
- CE 0344
- EMC: DIN EN 61000-4-2, DIN EN 61000-4-3, DIN EN 61000-4-4, DIN EN 61000-4-6 & DIN EN 55011
- Ex II 2 G
Approvals & Certifications
- Certificate of Compliance with the Order According to EN 10204: 2004- 2.1
- Flow Table for Recalculation
- Pressure Test Report for Measuring System
Consumption
- 2.2 mA According DIN EN 60947-5-6 (NAMUR) - Float Above Ring Sensor
Diameter
- Tube Diameter: 0.39" (10 mm)
- Tube Diameter: 0.67" (17 mm)
Differential Pressure
- > psi: 5.8 (kPa: 40, Mbar: 400), Maximum
Electrical Connectors
- 0.4 mm², 2 meters Long
- 2 x 0.14 mm² with Shield
Electrical Contacts
- Bistable Reed Contact
- Changeover Contact
Electrical Rating
- 250 Vac / 4 A / 500 VA, Maximum
Enclosure, Body Material
- Chromium Nickel-Steel - Controller & Springs
Environmental Protection
- Explosion-proof: Marking According to Guideline 94/9/EG, Ex Ia IIC 2 G T6
- Intrinsic Safe: [EEx Ia] II C According PTB 00-ATEX-2081 & PTB 00-ATEX-2080
- IP67
Flow Rate
- 0.002 to 110 L/hour to Water at 20° C (68° F)
- 0.1 to 3,500 L/hour to Air at 20° C (68° F), 1 bar Absolute
- 100 L/hour, Maximum
- 3,000 L/hour at 80° C (176° F) - Gas Flow (Controller)
Gasket Material
- Buna-N
- Polyethylene
- PTFE
- Viton
Length
- Tube Length Typical: 150 mm (5.90")
- Tube Length Typical: 300 mm (11.81")
- Tube Length Typical: 75 mm (2.95")
Materials of Construction
- 316Ti Stainless Steel - Fitting
- Corundum - Float Material
- CrNi-Ball - Float Material
- Glass Ball - Float Material
- MuMetal - Float Material
- Polypropylene - Fitting
- PTFE - Membrane
- PVDF - Limit Switch & Float Material
- Titanium - Float Material
Maximum System Pressure
- psi: 232.1 (kPa: 1,600, bar: 16)
Measurement Accuracy
- 1.6% According to Directive VDI/VDE 3513
- 2.5% According to Directive VDI/VDE 3513
- 4% According to Directive VDI/VDE 3513
- 6% According to Directive VDI/VDE 3513
Mounting
- Tapped Holes
NPT Connection
- 1/4" NPT
O-Ring Material
- Viton
Power Supply
- 115 Vac
- 230 Vac
- 24 Vdc
- 4.5-15 Vdc
Process Connection Material
- 316Ti Stainless Steel - Cutting Ring, Nozzle, Swagelok Connection, Inner Thread & Fitting Material
- Polypropylene - Inner Thread & Fitting Material
- Steel - Cutting Ring, Nozzle
Relay Output
- 1 or 2 Potential Free Changeover Contacts
Seat / Seal Material
- Valve Seat: Silver
Sensor, Probe Diameter
- 0.06" to 0.61" (1.6 to 15.7 mm)
Sensor, Probe Style
- Float
Spring Material
- Chromium Nickel-Steel
Switch Type
- Bistable Inductive Ring Sensor
- NAMUR
Temperature Range
- 100° C (212° F), Maximum
- 130° C (266° F), Maximum
- -25° to 65° C (-13° to 149° F)
- 80° C (176° F), Maximum