MCS300P Extractive Gas Analyzer

Brand: SICK Canada

The MCS300P is an extractive process photometer for measuring gaseous or liquid media. It measures IR and VIS active components with variable measuring ranges from very low (ppm) to high (vol%) concentrations. 

For monitoring of toxic or flammable mixtures, it has special process cells with safety devices like twinseals and flushing gas feeds. The heatable cells made of corrosion-resistant materials have a high pressure resistance. Automatic adjustment, innovative operation concept and modern communication protocols make the MCS300P an all-purpose photometer, even for potentially explosive atmospheres.

At a glance
  • Simultaneous measurement of up to 6 components
  • Process cells up to 60 bar and 200° C
  • Automated sample point switching
  • Integrated adjustment unit (optional)
  • Safety devices for measurement of toxic or flammable mixtures
  • Extended operation via PC and SOPAS ET software
  • Flexible I/O module system
Features
  • Automatic adjustment without expensive test gases
  • Integration into existing networks
  • Integration of external parameters like temperature or pressure
  • Suitable for explosive environments
  • Easy installation and maintenance thanks to transparent, compact design

Specifications

Certifications & Approvals

  • CE
  • ATEX
  • IECEx

Enclosure Rating

  • IP65

Measured Value

  • Br2, CCl4, C2Cl4, CCl2F2, C3F6, CH4, C2H2, C2H4, C2H6, C3H4, C3H6, C3H8, C4H6, C4H8, C4H10, C6H6, C6H14, C7H8, CH3Cl, CH2Cl2, CHCl3, C2H3Cl, C2H2Cl2, C2H4Cl2, C2HCl3, C6H5Cl, C6H4Cl2, CHCl2F, CH3OH, C3H7OH, C2H4(OH)2, CH3CHO, HCOOH, CH3COOC2H3, C2H2O, Cl2, CO, CO2, COCl2, HCl, HCN, HF, H2O, NH3, NO, N2O, NO2, R-NCO, SO2, SiF4, Many Other IR / VIS Active Gases and Liquids

Spectral Range

  • VIS Version: 300 to 1,200 nm
  • IR- Version: 1,200 to 12,000 nm

Measuring Range

  • More Than 60 Measuring Components Available (Depending on Concentration and Sample Gas Composition)
  • Up to 6 Components Simultanously
  • 2 Measuring Ranges Per Component
  • Automatic Measuring Range Switching (Adjustable)
  • 2 Limit Values Per Component
  • Measuring Ranges Depend on Application and Combination of Measuring Components

Response Time

  • 30 to 120 s

Ambient Temperature

  • 5° to 40° C Temperature Change Maximum ±10° C/h

Storage Temperature

  • -20° to 60° C

Humidity

  • ≤ 80% Non-condensing

Communication Interface

  • Modbus
  • Ethernet

Material of Construction

  • Aluminium, Coated

Power Supply

  • Standard Version 115 Vac, ± 10% / 230 Vac, ± 10%
  • Ex-version 230 Vac, ± 10%
  • Ex Control Device 230 Vac, ± 10%
  • Analyzer: 50 to 60 Hz, Ex Control Device: 48 to 62 Hz
  • Power Consumption: Analyzer: ≤ 230 VA;  With Cuvette Heating: ≤ 805 VA; With 2nd Heating: ≤ 1,450 VA
show more