Brand: Yokogawa

The WT5A / WT5V is a compact, front terminal connection type signal conditioner that is connected to an IEC/JIS-standard thermocouple (TC), such as a Type K, T, E, J, R, S, B or N thermocouples to convert temperature signals into isolated DC current or DC voltage signals.
Features
  • Selection of input type, input range setting, burnout setting, output adjustment, I/O monitoring,and loop back test can be made using the optional parameter setting tool (VJ77) or handy terminal (JHT200)
  • For the fahrenheit display, specify the option "/DF"
  • Available for the combination with safety barrier (BARD-600)
  • WT5A and WT5V are also available in 2,000 Vac voltage withstand specifications

Specifications

Accuracy

  • ± 0.1% of Span or ± 10 µV whichever is greater
  • ± 1° C (± 1.8° F) [0° to 50° C (32° to 122° F)] - Reference Junction Compensation for Thermocouple K, T, E, J, B, N
  • ± 2° C (± 3.6° F) [0° to 50° C (32° to 122° F)] - Reference Junction Compensation for Thermocouple R, S

Burn-Out

  • 60 seconds, Maximum
  • Down
  • Off
  • Up

Consumption

  • 45 mA at 24 Vdc (WT5V) - Power Consumption
  • 65 mA at 24 Vdc (WT5A) - Power Consumption

Current

  • 0.1 µA - Burn-out Detective Current

Dissipation

  • 3 VA at 100 Vac (WT5V) - Power Dissipation
  • 4 VA at 100 Vac (WT5A) - Power Dissipation
  • 4.5 VA at 200 Vac (WT5V) - Power Dissipation
  • 5.5 VA at 200 Vac (WT5A) - Power Dissipation

Electrical Connectors

  • M4 Screw Terminal

Humidity

  • 5 to 90% Relative Humidity, Non-Condensing - Operating Humidity

Insulation Resistance

  • 100 MegaOhms Minimum at 500 Vdc between Input, Output & Power Supply Mutually (DC-Driven), between Input, Output & Power Supply & Ground Mutually (AC-Driven)

Measuring Range

  • -270° to 1,820° C (-454° to 3,308° F) - Thermocouples

Measuring Span

  • 3 mV or more - Thermocouples

Mounting

  • DIN Rail Mounting
  • Rack-Mounted
  • Wall Mounted

Operating Temperature

  • 0° to 50° C (32° to 122° F)

Output

  • 0 to 24 mA DC
  • 10 mV-20 V - Span (DC)
  • -10-10 Vdc
  • 1-24 mA - Span (DC)

Power Supply

  • 100-240 Vac (-15%, 10%) (AC Drive)
  • 24 Vdc ± 10% (DC Drive)

Power Supply Effect

  • ± 0.1% of Span or less for Fluctuation within the Operating Range of Power Supply Voltage

Resistance

  • 1 KiloOhm or less - Signal Source Resistance
  • 1 MegaOhm or more - Input Resistance when Power On
  • 10 KiloOhms or more - Input Resistance when Power OFF
  • 2 to 250 KiloOhms or more - Allowable Load Resistance for DC Voltage
  • 750 to 15,000 ohms or less - Allowable Load Resistance for DC Current

Response Time

  • 200 milliseconds, 63% Response (10 to 90%)

Sensor, Probe Type

  • B Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • E Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • J Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • K Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • N Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • R Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • S Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)
  • T Thermocouple (ITS-90, JIS C 1602:'95, IEC 584:'95)

Span & Zero Adjustments

  • ± 1% of Span - Input Adjustment
  • ± 10% of Span - Output Adjustment
  • ± 5% of Span for -10-10 Vdc Output

Supply Voltage

  • -0.5 to 4 Vdc - Permissible Applied Voltage
  • 1,500 Vac for One Minute between Input, Output & Input, Power Supply - withstand Voltage
  • 500 Vac for One Minute between Output & Power Supply - withstand Voltage

Temperature Effect

  • ± 0.2% of Span or less for a Temperature Change of 10° C (18° F) - Effect of Ambient Temperature Change

Zero Offset

  • 0 to 200% - Zero Elevation (Current Signal)
  • -100 to 200% - Zero Elevation (Voltage Signal)
show more