Reliability and precision are the distinguishing features of the GME700 laser gas analyzer. Based on the principle of TDLS (Tunable Diode Laser Spectroscopy) and by using specific light absorption, the GME700 is able to measure several gas components, such as NH3, HF, or HCl. The compact 19" unit includes the analyzer and evaluation unit as well as a heated multipath measuring cell.

At a glance
  • High selectivity due to high spectral resolution
  • No calibration required
  • No moving parts: minimal wear and tear
  • Heated multipath measurement cell
  • Hot-wet measurement technology
Features
  • Optimal adaptation to measuring task through suitable gas conditioner
  • High reliability during operation
  • Can also be used in harsh ambient conditions
  • Easy integration into existing extractive systems
  • No problematic test gases required

Specifications

Certifications & Approvals

  • CE

Enclosure Rating

  • IP20

Measuring Range

  • HCl: 0 to 5 ppm / 0 to 3,000 ppm
  • HF: 0 to 5 ppm / 0 to 2,000 ppm
  • NH3: 0 to 25 ppm / 0 to 4,000 ppm
  • HCl / H2O: 0 to 5 ppm / 0 to 3,000 ppm (HCl), 0 to 25 Vol.% / 0 to 100 Vol.% (H2O)
  • NH3 / H2O: 0 to 25 ppm / 0 to 4,000 ppm (NH3) 0 to 3 Vol.% / 0 to 15 Vol.% (H2O)

Measured Value

  • HCl, HF, NH3, H2O

Response Time

  • ≤ 360 s

Accuracy

  • 2% to 4% of Measuring Range Full Scale

Sample Gas Temperature

  • ≤ +120° C Temperature at Analyzer Inlet

Process Pressure

  • 600 to 1,200 hPa

Ambient Temperature

  • 0° to 50° C

Storage Temperature

  • -40° to 55° C

Humidity

  • Ambient: ≤ 85% Non-condensing

Communication Interface

  • PROFIBUS DP
  • RS-232

Power Supply

  • 115 Vac / 230 Vac, 50 Hz / 60 Hz, Power Consumption ≤ 500 VA